Abstract

    Open Access Research Article Article ID: APRC-3-117

    Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP noninvasive Ventilation

    Marwa Mohsen, Ahmed A Elberry, Abeer Salah Eldin, Raghda RS Hussein and Mohamed EA Abdelrahim*

    Objective: Although, the use of humidification during non-invasive ventilation (NIV) is an important factor in decreasing nasal airway resistance and assuring patient’s comfort and adherence; many in-vitro studies recommend switching off the humidifier while delivering aerosol to NIV patients.

    The aim of the study was to in-vivo determine the effect of humidification on salbutamol delivered via different  inhalation  devices  to  chronic  obstructive  pulmonary  disease  (COPD)  patients  using  automatic continuous positive airway pressure (Auto-CPAP).

    Method: Aerosol delivery to NIV COPD patients by Aerogen Solo vibrating mesh nebuliser (VMN), Jet nebuliser (JET) and a metered dose inhaler (MDI) with AeroChamber MV spacer (AC) were compared with and without humidification. Auto-CPAP was adjusted at non-invasive ventilation mode with the integrated heated humidifier (IHH), as a source of humidity. The heater was set to the default setting 3, equivalent to 50ºC. Urine samples, 30 minutes and 24 hours post inhalation, as an index of the relative pulmonary and systemic bioavailability respectively, were provided by subjects and aliquots were retained for salbutamol analysis using solid phase extraction and high performance liquid chromatography (HPLC).

    Results: There  was  no  significant  difference  in  the  urinary  excreted  salbutamol  post  inhalation between the humidified and dry conditions. However, there was a significant difference between devices.The MDI with AC spacer had the highest percentage of 30 minutes urinary excreted salbutamol and JET had the lowest (p<0.01). VMN the highest percentage of 24 hours urinary excreted salbutamol and JET had the lowest but the difference was not significant.

    Conclusion: Significance of switching off humidity during aerosol delivered to ventilated patient was not as previously shown in in-vitro literatures. We recommend delivering aerosol medication to Auto-CPAPNIV patient using humidity without fear of lower delivery

    Keywords: Nebulizer; Non-invasive ventilation; Humidification; Urinary salbutamol; Filter; CPAP

    Published on: Mar 18, 2017 Pages: 11-15

    Full Text PDF Full Text HTML DOI: 10.17352/aprc.000017
    CrossMark Publons Harvard Library HOLLIS Search IT Semantic Scholar Get Citation Base Search Scilit OAI-PMH ResearchGate Academic Microsoft GrowKudos Universite de Paris UW Libraries SJSU King Library SJSU King Library NUS Library McGill DET KGL BIBLiOTEK JCU Discovery Universidad De Lima WorldCat VU on WorldCat

    Indexing/Archiving